Fig. 4From: AnemoCheck-LRS: an optimized, color-based point-of-care test to identify severe anemia in limited-resource settingsBland-Altman plots. The correlation between the AnemoCheck-LRS and the gold standard method was consistent across the spectrum of measured hemoglobin values (panel a). Most AnemoCheck-LRS results were within 0.5 (64%) or 1.0 (86%) g/dL of the measured Hb. In contrast (panel b), for the WHO Haemoglobin Color Scale (WHO HCS), the correlation appeared to worsen with increasing hemoglobin values. The WHO HCS consistently overestimated Hb values by as much as 8 g/dLBack to article page